An Absorbance of Two Means One Per Cent of the Light Gets Through
Gepubliceerd op 19-10-2026 · 4 min leestijd · Dagelijkse calculators
Lena Hoffmann — Redacteur wetenschap en onderwijs bij OneKitly
Wiskunde · Natuurkunde
Getoetst aan 2 bronnen
The Beer–Lambert law says absorbance is the molar absorptivity times the concentration times the path length. With a molar absorptivity of 6,220 M⁻¹cm⁻¹, a 100 µM solution and a standard 1 cm cuvette, A comes to 0.622. What that number means is easy to misread: absorbance is a base-ten logarithm of the fraction of light blocked, so A = 0.622 lets 23.9 per cent of the beam through, A = 1 lets 10 per cent, A = 2 lets 1 per cent and A = 3 lets a thousandth. The relationship to concentration is perfectly linear — halving the sample halves A to 0.311, doubling the cuvette doubles it to 1.244 — but the relationship to the light reaching the detector is not, and that is where the measurement runs out.
6,220 M⁻¹cm⁻¹ at 100 µM through 1 cm gives A = 0.622, and 23.9 % of the light emerges. Double the path length and A doubles to 1.244 — the scale is logarithmic, and that is where it breaks.
Dilute until the number lands between 0.1 and 1
At A = 3 the detector is comparing a beam to a thousandth of itself, and every source of stray light, every scratch on the cuvette and every particle of dust contributes a larger share of what remains than the sample does. At A = 0.05 the opposite problem appears: the difference between sample and blank is smaller than the instrument's own drift. The usable band in between is where a spectrophotometer earns its keep, and getting into it is a matter of choosing the dilution or the cuvette — a 1 mm path instead of 10 mm divides the reading by ten without touching the sample at all, which is the fastest fix when material is scarce.
The absorptivity belongs to a compound at a wavelength
It is not a property of the machine or of the sample tube; it is a number for one substance measured at one wavelength in one solvent, and moving any of those three changes it. Quoting an absorptivity without the wavelength is like quoting a speed without a direction — the figure is real and the sentence is incomplete. It is also why a spectrophotometer is set to a specific wavelength before anything is measured, usually the peak of the compound's own absorption, where the reading is largest and least sensitive to the instrument being slightly off. A shift of a few nanometres near a steep flank of the spectrum can move the answer by more than the whole dilution error.
| Absorbance | Light transmitted | Verdict |
|---|---|---|
| 0.1 | 79.4 % | Usable, near the noise floor |
| 0.622 | 23.9 % | Comfortable |
| 1 | 10 % | Top of the good range |
| 2 | 1 % | Dilute the sample |
| 3 | 0.1 % | Not a measurement |
Uitgewerkt met onze eigen calculator
Beer-Lambert-calculator (wet van Beer-Lambert)
Gegeven
- Los op naar
- Concentratie (c)
- Absorbantie A (indien bekend)
- 1,244
- Molaire extinctiecoëfficiënt ε (M⁻¹cm⁻¹)
- 12.440
- Concentratie c
- 200
- Eenheid concentratie
- mM
- Weglengte l (cm)
- 2
Resultaat
- Resultaat
- 0,05
Deze cijfers komen uit de calculator hieronder, ze zijn niet met de hand ingetypt — ze worden opnieuw berekend telkens als de tool verandert.
Reken het door met je eigen cijfers →Veelgestelde vragen
- What is the blank actually for?
- To subtract everything that is not the thing you are measuring. The solvent absorbs a little, the cuvette walls reflect a little, and the lamp is not equally bright at every wavelength; zeroing on a cuvette containing everything except the analyte removes all three in one step. A blank made from a different batch of solvent, or run in a different cuvette from the sample, quietly reintroduces what it was supposed to remove — which is why paired cuvettes exist and why the blank is remeasured whenever the wavelength changes.
- Why does the law fail at high concentration?
- For two reasons at once, one about the instrument and one about the chemistry. The instrumental one is stray light, which sets a ceiling on any reading: once the true transmission falls below the light leaking around the sample, further concentration changes nothing. The chemical one is that crowded molecules stop behaving independently — they interact, associate, and shift each other's absorption — so the absorptivity that was constant at micromolar concentrations is no longer the same number at millimolar. Both push the calibration curve to bend downward, which is why a standard curve is run across the range actually used rather than extrapolated from one point.